We are able to support new customers with cables, chargers and sensors. Advanced Myographic Technologies keeps a stock so they can overnight supplies as necessary. CURO Diagnostics and AMT both provide support via emails. Contact us at (mail@myographytech.com) or you may call us at (352-456-8338) or visit our website at www.myographytech.com/shop/ as necessary.
If you experience a lack of connection between the CURO and your iPad then the most likely reason is an incorrect WiFi setting.
Please check that the iPad WiFi connection is set up correctly .. it should appear as "amg" on the iPad when the CURO second "blue" button is activated .. then check that the setup for the "amg" WiFi is as follows:
IP Address - 192.168.1.9
Subnet Mask - 255.255.255.0
Router - 192.168.1.1
When you use a CURO-Diagnostics AMG sensor and you change the level of amplification in the APP from 0dB gain all the way up to 32dB gain, the maximum amplitude of the recorded signal declines in an exponential way - see the figure below for details or use the equation
y = 2.2114e-0.086x where x = dB and y is the amplitude in Volts.